Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates

M. Z. Arsalan, M. B. Kakakhel, M. Shamshad, T. A. Afridi

Abstract


Continuous gantry motion, continuous beam modulation, and variable dose rate are used in volumetric modulated arc therapy (VMAT) to obtain highly conformal radiation therapy dose distributions. Several errors during daily radiation therapy treatment can be sources of uncertainties in dose delivery. These errors include monitor unit calculation errors and other human mistakes. Due to the uncertainties in the excessively modulated VMAT plan, the intended dose distribution is not delivered perfectly, leading to a mismatch between the measured and planned dose distributions. This necessitates an extensive and effective quality assurance (QA) program for both machine and patient. In this study, VMAT QA plan verification of 62 head and neck (H&N) and 19 prostate cases was done using Octavius 4D setup with its associating VeriSoft gamma analysis software. The plans showed a maximum 3D gamma passing rate with 4 mm/3 % gamma acceptance criteria, i.e., 99.7 % for the H&N cancer cases and 99.5 % for the prostate cancer cases. Local gamma analysis was also performed for both regions. Furthermore, 2D and volumetric gamma analyses were also carried out. Gamma analysis with respect to different axis was also carried out. It was known that the transversal axis showed    the highest gamma passing rate in both H&N and prostate cases, i.e., 99.17 % and 98.3 %, respectively. The transverse axis came to be a better fit for the planned dose distribution.


Keywords


VMAT, Treatment Planning System Gamma Analysis Octavius 4D Setup Gamma Acceptance Criteria

Full Text:

PDF

References


E. Kamperis, C. Kodona, K. Hatziioannou et al., Int. J. Radiat. Oncol. Biol. Phys. 106 (2020) 182.

F. M. Khan, P. W. Sperduto and J. P. Gibbons, Khan’s Treatment Planning in Radiation Oncology: Lippincott Williams & Wilkins, (2021).

M. Antoine, F. Ralite, C. Soustiel et al., Phys. Medica 64 (2019) 98.

D. P. Goodwin, Point dose measurements in VMAT: an investigation of detector choice and plan complexity. M. Sc. Thesis, University of Canterbury (2017).

R. Wang, Y. Du, K. Yao et al., Phys. Medica 71 (2020) 14.

Y. Wang, X. Pang, L. Feng et al., Phys. Medica 47 (2018) 112.

T. Milan, G. Grogan, M.A. Ebert et al., Med. Phys. 46 (2019) 1984.

R. Holla, V.A. Ravikumar and B. Pillai, Med. Phys. 40 (2013) 368.

S. Michiels, K. Poels, W. Crijns et al., Radiother. Oncol. 128 (2018) 479.

M. Hussein, P. Rowshanfarzad, M.A. Ebert et al., Radiother. Oncol. 109 (2013) 370.

R. Mahmood, A. Babier, A. McNiven et al., Automated Treatment Planning in Radiation Therapy using Generative Adversarial Networks, Proceeding of the 3rd Machine Learning for Healthcare Conference, PMLR 85 (2018) 484.

P. D. H. Wall and J. D. Fontenot, Phys. Medica 87 (2021) 136.

S. C. Peet, L. Yu, S. Maxwell et al., Phys. Medica 78 (2020) 166.

S. Diamantopoulos, K. Platoni, G. Patatoukas et al., Phys. Medica 67 (2019) 107.

V. Wyrwoll, B. Delfs, M. Lapp et al., Nucl. Instrum. Methods Phys. Res. Sect. A 987 (2021) 164831.

B. Pal, A. Pal, S. Bag et al., Phys. Medica 91 (2021) 18.

A. G. Ibrahim, I.E. Mohamed and H.M. Zidan, Int. J. Med. Phys., Clin. Eng. Radiat. Oncol. 7 (2018) 438.

D. Mönnich, J. Winter, M. Nachbar et al., Phys. Med. Biol. 65 (2020) 1.

C. Q. Garcia, E. C. Solis, J. V. Salazar et al., Rev. Mex. Fis. 67 (2021) 061101.

E. Decabooter, A. C. C. Swinnen, M. C. Öllers et al., Br. J. Radiol. 94 (2021) 20210473.

B. Chand, M. Kumar and M. Kumar, Eur. J. Mol. Clin. Med. 7 (2020) 3595.

J. M. Park, J. Kim, S.-Y. Park et al., Radiat. Oncol. 13 (2018) 1.

I. A. P. I. Gayatri, A. D. Handika, W. E. Wibowo et al., Appl. Radiat. Isot. 188 (2022) 110415.

M. El Kafhali, M. Khalis, M. Tahmasbi et al., Phys. Med. 12 (2021) 100044.

A. Chen, J. Zhu, N. Wang et al., Phys. Medica 90 (2021) 134.

S. Zouiri, M. Tantaoui, M. El Hassani et al., Validation of Monaco Treatment Planning System for Intensity Modulated Radiation Therapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT®) on ELEKTA InfinityTM Linear Accelerator, E3S Web Conferences 297 (2021) 01014.

M. Chun, H. J. An, O. Kwon et al., Phys. Medica 62 (2019) 83.

U. Yoshihiro, O. Shingo, I. Masaru et al., Br. J. Radiol. 91 (2018) 1.

Y. C. Lee and Y. Kim, Int. J. Med. Phys., Clin. Eng. Radiat. Oncol. 10 (2021) 16.

A. Bruschi, M. Esposito, S. Pini et al., Phys. Medica 49 (2018) 129.

W. Woon, P. B. Ravindran, P. Ekayanake et al., J. Appl. Clin. Med. Phys. 19 (2018) 230.




DOI: https://doi.org/10.55981/aij.2023.1261



Copyright (c) 2023

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.