Long-Term Stability and Performance Assessment of 137Cs Gamma Beams for Calibration of Radiation Protection Instruments According to ISO 4037 and IAEA Standards
Abstract
This study presents the experimental characterization of gamma radiation beams generated by a collimated irradiator equipped with two ¹³⁷Cs sources at the Gamma and X-ray Calibration Laboratory of CNESTEN. The facility is mainly used for calibrating radiation protection instruments and supporting dosimetric assessments. The irradiator provides air kerma rates (Kair) from 60 µGy/h to 60 mGy/h at distances up to three meters. Calibrations were performed in terms of ambient dose equivalent using Kair measurements and appropriate conversion coefficients, with corrections applied for source decay to maintain accuracy over time. A long-term evaluation of the irradiator’s stability over 16 years showed consistent performance, with Kair rate deviations within 5% after decay correction. This confirms the reliability of the reference beams and supports extending calibration intervals. To ensure calibration validity, a beam homogeneity study was also conducted by measuring the transversal profile at various distances from the source. The results showed a uniform beam width of approximately ±8 cm at 1 meter and up to ±21 cm at 3 meters from the central axis, confirming the homogeneity of the irradiator for precise and reliable measurements across different distances. Finally, the system’s performance meets the requirements of ISO 4037:2019 and IAEA SRS-16 standards. The expanded uncertainty of ±3.7% (k = 2) reflects the high reliability of dosimetric evaluations at the laboratory.
Keywords
Full Text:
PDFReferences
E. B. Jumpeno, R. Anggraeni, S. N. Barokah et al., Atom Indones. 49 (2023) 209.
T. Zidouz, E. M. Alibrahmi, E. M. Chakir et al., Int. J. Power Energy Convers. 14 (2023) 332.
T. Zidouz, A. Talbi, E. M. Chakir et al., Appl. Radiat. Isot. 215 (2025) 111588.
International Atomic Energy Agency (IAEA), Calibration of Radiation Protection Monitoring Instruments, Safety Reports Series No. 16, IAEA, Vienna (2000).
W. Purwanti, F. Suhaimi, W. E. Wibowo et al, Atom Indones. 49 (2023) 27.
Y. Lusiyanti, Z. Alatas, M. Lubis et al, Atom Indones. 39 (2013) 124.
ISO 4037-1:2019, Radiological Protection - X and Gamma Reference Radiation for Calibrating Dosemeters and Doserate Meters and for Determining Their Response as A Function of Photon Energy, Part 1: Radiation Characteristics and Production Methods, International Organization for Standardization (ISO), Geneva (2019).
E. Fernandes, D. Freire, A. C. de Freitas et al., Appl. Radiat. Isot. 61 (2004) 1425.
C. Hranitzky and H. Stadtmann, Nuclear Nucl. Instrum. Methods Phys. Res. A 580 (2007) 230.
Nucomat, Equipment Manual: Gamma Collimated Irradiator Morocco (Canberra Eurisys Ref. VMVH/02026), Nucomat (2002).
ISO 4037-2:2019, Radiological Protection - X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as A Function of Photon Energy, Part 2: Dosimetry for radiation Protection Over the Energy Ranges from 8 keV to 1.3 MeV and 4 MeV to 9 MeV, International Organization for Standardization (ISO), Geneva (2019).
ISO 4037-3:2019, Radiological Protection - X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as A Function of Photon Energy, Part 3: Calibration of Area and Personal Dosemeters and The Measurement of Their Response as A Function of Energy and Angle of Incidence, International Organization for Standardization (ISO), Geneva (2019).
Suliman, B. E. Youssif, A. A. Beineen et al., Radiat. Meas. 45 (2009) 1578.
J. Gräfe, Y. Poirier, F. Jacso et al., J. Appl. Clin. Med. Phys. 15 (2014) 356.
Joint Committee for Guides in Metrology, Evaluation of Measurement Data - Guide to The Expression of Uncertainty in Measurement (GUM), (JCGM 100:2008), JCGM, Sèvres (2008).
DOI: https://doi.org/10.55981/aij.2026.1673
Copyright (c) 2026 Atom Indonesia

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.